Portable Gas Detectors covers the personal and transportable monitors that warn the wearer of flammable gas, hydrogen sulphide, carbon monoxide and oxygen deficiency, plus their docking stations and sampling pumps.

Each monitor is an intrinsically safe instrument with consumable sensors and a rechargeable battery. Its remaining sensor life, battery health and calibration history matter more than its age.

What a personal monitor does

A personal monitor samples the air at the wearer's breathing zone and alarms by sound, light and vibration when a reading passes a set point. The common four-gas monitor reads flammable gas in %LEL, oxygen in % volume, and H2S and carbon monoxide in ppm. Single-gas H2S monitors are worn by everyone near the well on sour wells. Five and six-gas models add sensors such as a photoionisation detector (PID) for volatile organic compounds or a second toxic sensor.

Configurations

TypeUseNote
Single-gas, disposableH2S or CO worn all shiftFixed life counted from activation; no recalibration or new sensor
Single-gas, serviceableH2S, CO or O2Sensor and battery replaceable
Multi-gas, diffusionClip-on monitor for general workReads only where it is worn
Multi-gas, pumpedPre-entry testing of tanks and pitsDraws a sample through a tube
Transportable area monitorTemporary perimeter during hot work or well testingSome link wirelessly to a controller

Key specifications

ParameterWhy it mattersTypical classes or units
Sensor setMust match the gases on your risk assessment%LEL, % vol O2, ppm H2S and CO
LEL sensor typeCatalytic needs oxygen; IR cannot see hydrogenCatalytic bead or infrared
Ex markingDecides which zones it may enterEx ia IIC T4, EPL Ga for Zone 0
Alarm set pointsTWA and STEL alarms for toxic gasesppm, set to your exposure limits
Data loggingProves exposure and bump test historyEvent and datalog memory

Standards and certification

The ISEA statement on validating portable gas monitors recommends a bump test or calibration check before each day's use, and a full calibration if it fails. Most makers follow that rule in their manuals.

Failure modes and inspection

On a used monitor, check the sensor dates and the event log first, then run a full calibration with certified gas and a bump test of every alarm.

Documentation

Integration with your fleet

Monitors from the same maker as your fleet share docking stations, calibration gas mixtures, chargers and spare sensors. A docking station automates bump tests and calibration and stores the records an auditor asks for. Check that the gas mixture the station needs matches the cylinders you already stock.

Refurbishment and what never to buy used

A serviceable monitor can be refurbished with new sensors, a new battery pack from the maker, new filters and a full calibration. Never buy a used disposable single-gas monitor, because its remaining life cannot be extended or verified. Avoid any unit with a missing date code or a battery of unknown origin, and expired calibration gas.

For fixed detection in the same areas, see gas and H2S detection. For respiratory protection during entry, see breathing air systems.

FAQ

Frequently asked questions

How often should a portable gas detector be bump tested?

Bump test it before each day's use, as the ISEA statement on portable gas monitors recommends and most makers require. If it fails, run a full calibration before use. A docking station can do both and keep the records.

Can a used disposable H2S monitor be reused?

No, a disposable single-gas monitor has a fixed life counted from activation and cannot be recalibrated or re-sensored. Buy these only new and sealed, with the activation deadline printed on the pack in date.

Can I fit a cheaper battery to a portable gas monitor?

No, fit only the battery pack the Ex certificate lists. A different pack voids the intrinsic safety certification and may overheat in a hazardous area. Check the pack's date code on a used monitor as well.

Why does my LEL reading drop in a confined space?

A catalytic bead LEL sensor needs oxygen to burn the gas, so it reads low when oxygen is deficient. Always read the oxygen value first. An infrared LEL sensor avoids this but cannot detect hydrogen.

What should I check on a used multi-gas monitor?

Check the date code of every sensor and the battery pack first. Then download the event log, inspect the housing and gaskets, and run a full calibration and bump test. Confirm the firmware works with your docking stations.